Jkyp-1 liquid chromatograph
Number of views:461
QR code:

Add time:2021-01-21
Brief description:
Main performance characteristics:
1. The pump head has a self flushing function, allowing for infusion without the need for manual emptying;
2. Adopting the "floating pump plunger rod sealing ring" technology can extend the life of the sealing ring and prevent leakage for 5000 hours;
3. Programmed solvent compression factor that can automatically compensate;
4. Complete pump types: analytical, semi prepared, and prepared;
5. The UV visible light detector adopts a holographic concave grating and dual optical path design, and the detection wavelength can be extended to the visible spectrum range;
6. Easy to replace the flow cell, convenient for daily maintenance;
7. The microcontroller control technology enables the instrument to have high automation and self checking functions, and all parameters are operated by the keyboard; 8. The circuit design, combined with excellent optical path structure, enables the detector to have low noise and baseline stability;
9. Automatically calculate the cumulative ignition time and number of ignitions of tritium lamps;
10. All parameters can be protected against power failure;
11. Adopting the 7725i injection valve from Rheodyne Corporation in the United States to ensure high reliability and reproducibility of the instrument;
12. Building block structure design, easy to expand into gradient elution system.
Main technical indicators:
High pressure constant current infusion pump:
1. Flow range: 0.001ml/min to 10.000ml/min (step size set to 0.001ml/min)
2. Flow accuracy: ≤± 1% (1.000ml/min, 8.5MPa, water, room temperature)
3. Traffic stability: RSD ≤ 0.25%
4. Pressure resistance: ≤ 40Mpa (0.001ml/min~10.000ml/min)
5. Pressure pulsation: < 0.1MPa (flow rate 1mL/min, pressure 5MPa~10MPa)
6. Power supply: AC 110V/220V 50Hz/60Hz)
7. Power consumption: 100W
Variable wavelength UV visible light detector:
1. Wavelength range: 190nm to 700nm
2. Wavelength repeatability: ± 0.01nm
3. Wavelength accuracy: ± 1nm
4. Band width: 6nm
5. Baseline noise: ≤ 2 × 10-5AU (methanol, 1ml/min, 254nm, 20 ℃)
6. Baseline drift: ≤ 4 × 10-4AU/hr (methanol, 1ml/min, stabilization time 40min, 254nm, 20 ℃)
7. Baseline noise: ≤ 0.5 × 10-5AU (empty pool, response time of 1 second, 20 ℃)
8. Baseline drift: ≤ 0.5 × 10-5AU/hr (empty pool, response time of 1 second, stabilization time of 40 minutes, 20 ℃)
9. Detection concentration: 6.0 × 10-9g/ml (naphthalene/methanol solution)
10. Flow pool volume: 10ul
11. Optical path length: 5mm
12. Power supply: AC 110V/220V 50Hz/60Hz)
13. Power consumption: 100W