JKZH60 SF6 gas integrated tester
Number of views:585
QR code:

Add time:2021-11-04
Brief description:
1、 Main functions and features:
1. Equipped with self diagnostic function
2. Cooperate with the upper computer software to upload data and output measurement values in EXCEL format
3. Equipped with temperature compensation and detection endpoint judgment
2、 Main technical indicators:
Humidity parameters | Range: -80~+20 ℃ Accuracy: ± 0.4 ℃, refer to some parameters of HNP-401FD |
Purity parameters | Range: 0-99% Accuracy: ± 0.6%, refer to some parameters of HNPSP-10 |
Decomposition product parameters | SO2:0~100μL/L,H2S:0~100μL/L,CO:0~1000μL/L,HF:0~10μL/L( Optional) Refer to the partial parameters of HNP-2011FA decomposition products |
Display | 800 × 480TFT touch screen |
Function combination | Dew point ppmv、 Purity H2S、SO2、CO、HF、 Flow rate, pressure, temperature |
Working Voltage | Built in lithium battery for both AC and DC use |
Work pressure | <1Mpa |
Sample gas flow rate | About 900mL/min |
Output interface | USB2.0 |
Storage function | 1000 items |
Printer selection | Thermal or needle type |
Operating environment | Environmental temperature: -20 to+50 ℃ |
Environmental humidity:<80% RH | |
Weight | About 8kg |
Size | About 400mm × 370mm × 180mm |
Usage time | About 8 hours |


SERVICE HOTLINE 